User Notes:
- Open the LM-300 LED Radiometer battery cover to connect the positive and negative
leads to the battery. Close battery cover. - Turn on the power switch located under the LCD screen. The screen will illuminate
Indicating the LM-300 LED Radiometer is ready for use. - Centre the tip of the light guide (8mm or larger) over the aperture of the LM-300 LED
Radiometer then activate the curing light directly the light source into the LM300 LED
Radiometer. Be sure to align the curing light tip directly over the aperture minimize the
the amount of escapable light. Read the output in mW/cm2 and note the reading for
comparison with the next routine test.
Less than 300 mW/cm2: Low output
300-500 mW/cm2: Test composite to determine curing time
500-800 mW/cm2: Use manufacturers recommended curing time
Over 800 mW/cm2: High output- reduce curing time appropriately
CAUTION: Equipment is not suitable for use in the presence of an anaesthetic mixture with air or nitrous oxide.
Specification:
- Atmospheric Pressure: 0.5 atm to 1.0 atm (500 hpa to 1060 hpa)
- Sensitivity Range: 400-500nm
- Intensity Range: 0-2,500 mw/cm2
- Accuracy: +- 5% of reading
- Overall Dimensions: 1.30” H x 4.53 “ L x 2.74” W
- Weight: 4.57 oz